• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTÉGÉ MRI IPG; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTÉGÉ MRI IPG; SCS IPG Back to Search Results
Model Number 3771
Device Problem Charging Problem (2892)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
It was reported the patient's programmer reflected an "ipg battery low" warning even though the ipg was charged.It was noted a spare charger also reflected the same warning.In addition, the patient's charger had never indicated a full charge since the patient was implanted.A replacement charger was sent to the patient which did not resolve the issue.Followup information revealed the patient underwent surgical intervention where the ipg was explanted and replaced.The patient reported effective therapy postoperative.The event date is unknown.
 
Event Description
Follow-up information revealed the patient is doing well postoperative.Surgical intervention resolved the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTÉGÉ MRI IPG
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
jennifer shepard
6901 preston road
plano, TX 75024
9725264657
MDR Report Key5508956
MDR Text Key40631695
Report Number3006705815-2016-00114
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2017
Device Model Number3771
Device Lot Number5310715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/25/2016
Initial Date FDA Received03/17/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-