Model Number 16413
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.The field service representative (fsr) verified that the arterial monitor on start up would reboot several times and then the display would go blank.During troubleshooting he found that the arterial monitor computer board was the source of the issue.The fsr replaced the arterial monitor computer board and checked arterial monitor calibrations and operation.The unit operated to manufacturer specifications and was returned to clinical use.The suspect part was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, when the arterial monitor powered up, it constantly rebooted then went blank.No audible alarms were heard.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) installed the returned board into a lab use only cardioplegia monitor.When power was supplied, the monitor¿s display would illuminate for a second and then not illuminate.Occasionally, the monitor would reboot but then the display would not illuminate.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 PERFUSION SYSTEM
Type of Device
MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS-SARNS 8000
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5513567
MDR Text Key41030092
Report Number1828100-2016-00198
Device Sequence Number1
Product Code DTW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16413
Device Catalogue Number16413
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received04/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number1828100-06/14/12-017-C
Patient Sequence Number1
-
-