The freedom driver was not supporting a patient.The customer reported that the freedom driver's display did not activate when the hospital staff member performed a system check.The freedom driver was returned to syncardia for evaluation.Visual inspection of the driver's external components revealed no abnormalities.Visual inspection of the driver's internal components revealed that the black wire in the airflow sensor cable was severed.The black wire in the airflow sensor cable connects the flow output signal from the flow sensor to the main printed circuit board assembly (pcba).It is unknown how the black wire was severed.The electronic record did not reveal any new fault alarms when the driver last left syncardia.Only permanent fault alarms are recorded in the driver's alarm history.Intermittent, recoverable and battery alarms are not recorded in the electronic record.The driver in "as received" condition failed testing as a result of the broken black wire in the air flow sensor cable, which resulted in a fault alarm as expected.The broken wire prevented the electronics of the driver from measuring the flow sensor reading and from displaying the appropriate calculations for fill volume (fv) and cardiac output (co).Therefore, as designed, asterisks are exhibited on the display when the values are out of range.The customer reported issue was consistently seen throughout the duration of the functional testing of the driver.A fault alarm was induced at approximately 15 minutes upon startup because that is when the low cardiac output alarm is enabled.The electronics displays the asterisks as a condition in which the cardiac output is less than 3.5 liters per minute (lpm), resulting in a low cardiac output alarm and a fault code of 49 recorded in the electronic record, which occurred during investigation testing.The customer-reported issue that the display would not activate was not reproduced.However, it is likely that the customer experience was referring to the displayed asterisks, which was caused by the damaged airflow sensor cable.Despite the customer-reported display issue, risk to the patient was low because the driver was not supporting a patient at the time, and this issue would not prevent the driver from performing its life-sustaining-functions.The driver was serviced, which included the replacement of the airflow sensor cable, before being placed into finished goods.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|