• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to zoll medical and the malfunction was verified.Visual examination found the main knob index ring installed incorrectly.There is no history of the device being serviced at zoll, therefore the root cause is customer tampering with the device.Main knob index ring was reinstalled to remedy the problem.The device was recertified and returned to the customer.No trend is associated with reports of this type.
 
Event Description
Complainant alleged that during functional testing, the device was unable to select the pace function with the main knob.Complainant did not indicate that there was any patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5523149
MDR Text Key41246728
Report Number1220908-2016-00695
Device Sequence Number1
Product Code DRO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060559
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 03/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Other Device ID Number00847946002695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/09/2016
Initial Date FDA Received03/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-