Catalog Number FEM10120
Device Problems Positioning Failure (1158); Premature Activation (1484)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Medwatch report # 9681442-2016-00091.
 
Event Description
It was reported that the endovascular stent graft partially deployed.Another stent graft was used to complete the procedure successfully.No additional treatment was required.No patient injury was reported.This is the same patient as reported in medwatch report # 9681442-2016-00091.
 
Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.
 
Event Description
It was reported that the endovascular stent graft partially deployed.Another stent graft was used to complete the procedure successfully.No additional treatment was required.No patient injury was reported.This is the same patient as reported in medwatch report # 9681442-2016-00090.
 
Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.As part of all complaint investigations, an attempt was made to evaluate the subject device as well as the device usage.In this case, no sample or image was provided for evaluation.Therefore, the reported event could not be reproduced.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent deployment failure.Insufficient flushing of the device may be another contributing factor to the reported event.On the basis of the information available and as no sample was returned, a definite root cause for the reported event could not be determined.The ifu states that the delivery system must be flushed with sterile saline.Furthermore, the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." also the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG -9681442
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
daniela mueller
wachhausstrasse 6
karlsruhe 76227
GM   76227
0497219445
MDR Report Key5533678
MDR Text Key41728773
Report Number9681442-2016-00090
Device Sequence Number1
Product Code PFV
UDI-Device Identifier04049519008622
UDI-Public(01)04049519008622(17)170801(10)ANYH0177
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130029/S003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2017
Device Catalogue NumberFEM10120
Device Lot NumberANYH0177
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-