Device Problem Improper Device Output (2953)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  No Answer Provided  
Event Description
A customer in (b)(6) reported their thinprep 5000 processor with autoloader etched a wrong vial id onto a slide.A customer reported intermittent wrong characters etched on the slides.This issue only occurred 3 or 4 times in the last month.Customer reported that samples were reprocessed, no impact to patient or delay in diagnosis reported.Customer sent pictures to hologic for further investigation.Hologic's field service engineer confirmed but unable to reproduce error.Performed 2d barcode camera upgrade ((b)(4)).Although no patients needed to be recalled as a result of this incident this is a reportable event since the instrument did not perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THINPREP 5000 PROCESSOR
Type of Device
THINPREP 5000 PROCESSOR
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer Contact
eva maxwell
250 campus drive
marlborough, MA 01752
MDR Report Key5536262
MDR Text Key42946611
Report Number1222780-2016-00074
Device Sequence Number1
Product Code MKQ
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 03/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/01/2016
Initial Date FDA Received03/30/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-