• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 7078396
Device Problem Metal Shedding Debris (1804)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/08/2016
Event Type  malfunction  
Manufacturer Narrative
The product is not returned to manufacturer for evaluation therefore cause of event cannot be determined.
 
Event Description
It was reported that on (b)(6) 2016 patient with adolescent idiopathic scoliosis underwent surgery at th10/l3.Intra-op, when the surgeon tried to insert a set screw after rotation correction, it was difficult to insert so the surgeon pushed it stronger.When the surgeon took off the ari from the set screw they were found to be cross-threaded and burrs were generated.Burrs were removed.The set screw was inserted with cross thread.No patient complications were reported.The procedure time was delayed over one hour.
 
Manufacturer Narrative
Product analysis :macroscopic and optical examination revealed thread crest and flank damage; this damage appears to have initiated at the start of the thread, and is consistent around the damaged portion of the thread.Functional evaluation with a sample mas found the implant unable to be fully engaged into the mas head.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
greg anglin
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key5537410
MDR Text Key41521382
Report Number1030489-2016-00864
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/23/2023
Device Catalogue Number7078396
Device Lot NumberH5196604
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-