Received one 12.5f x 28cm hemo-cath for evaluation.A visual inspection reveals the catheter to be in 2 pieces.There is a small tear/cut in the lumen right below the hub.A review of the manufacturing records indicated that all device specifications and quality requirements were satisfied.This family of devices is 100% leak tested during the manufacturing process and inspection.There is no evidence of swelling, crumbling, discoloration, degradation or polymer variation of the lumen material.The catheter was inserted and used for 2.5 years prior to the failure.There is no evidence of a manufacturing defect.We are unable to determine the cause or factors that may have contributed to this event.It appears that the catheter may have come into contact with a sharp object.
|