• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number MICROBLENDER
Device Problem Improper Device Output (2953)
Patient Problem No Patient Involvement (2645)
Event Date 03/09/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).In the event the additional information is received, a follow-up report will be submitted.Results of investigation: the carefusion factory service technician was able to duplicate the reported issue and confirmed that the oxygen was being delivered at 81 percent when it was set to 90 percent.The root cause was due to the blender being out of specification.There unit received overhaul service, was tested, and it meets all service specifications.(b)(4).
 
Event Description
The customer stated that the microblender fails the oxygen test.The customer requested that the blender be sent in for repair.There was no patient involvement at the time reported incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROBLENDER
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
7149227830
MDR Report Key5553179
MDR Text Key42076774
Report Number2021710-2016-03472
Device Sequence Number1
Product Code BZR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMICROBLENDER
Device Catalogue Number03800F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-