Model Number H965SCH647120 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: (b)(6) or older.(b)(4).
|
|
Event Description
|
It was reported that stent deployment issue occurred.The target lesion was located in the carotid artery.A 10.0-24 carotid wallstent¿ was advanced to treat the lesion.However, the stent was not fully deployed to the vessel wall.The procedure was completed with a different device.No patient complications were reported and the patient's status was good.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: a visual examination of the returned device identified no kinks or damage which could potentially have contributed to the complaint incident.A recommended 0.014 inch guidewire was inserted through the wire lumen with no resistance noted.An examination of the inner and outer exit holes found no damage.As part of the device analysis the stent was fully deployed with no restriction.An examination of the deployed stent found no damage.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Event Description
|
It was reported that stent deployment issue occurred.The target lesion was located in the carotid artery.A 10.0-24 carotid wallstent was advanced to treat the lesion.However, the stent was not fully deployed to the vessel wall.The procedure was completed with a different device.No patient complications were reported and the patient's status was good.
|
|
Search Alerts/Recalls
|