MAUDE Adverse Event Report: MPRI ATTAIN ABILITY PLUS; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN ABILITY PLUS; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 429688
Device Problems Failure to Capture (1081); High Capture Threshold (3266); Unstable Capture Threshold (3269)
Patient Problem Syncope (1610)
Event Date 12/07/2015
Event Type  Injury  
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.Product event summary:the lead was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated pacing capture threshold in the rv (right ventricle) was elevated.Right ventricular capture threshold is variable butis generally greater than 2v from (b)(6) 2014 to (b)(6) 2016.(b)(4).
 
Event Description
It was reported that the right ventricular (rv) lead and left ventricular (lv) lead had high and variable thresholds with loss of capture after the patient had a fall.The patient also complained of syncope when laying on their side the leads were reprogrammed and remain in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY PLUS
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5557389
MDR Text Key42231436
Report Number2649622-2016-03252
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/10/2014
Device Model Number429688
Device Catalogue Number429688
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age00075 YR
-
-