(b)(4).As returned the threaded pin is seized in the humeral cut guide.Dimensional analysis of the guide found the remaining holes conforming.Dimensional analysis of the pin also found it conforming where measured.Additionally, galling and wear can be seen on the portion of the pin protruding from the guide.Review of the receiving inspection documents for the guide did not find any deviations or anomalies.The devices are used for treatment.Based on its lot number, the guide has an approximate field age of 7 years, 6 months.It is possible that damage to the guide and/or the pin was accrued during previous use and this damage caused interference between the two.This interference could have generated particulates that became trapped between the pin and the guide, thus potentially causing friction and eventually device seizure.The instructions for the guide states "where instruments form part of a larger assembly, check that the devices assemble readily with mating components.If damage or wear is noted that may compromise the function of the instrument, do not use the device and contact your zimmer representative for a replacement." the previously accrued damage can most likely be attributed to normal wear and tear.This mdr was identified during an internal retrospective review to meet reporting requirements and therefore is being filed retrospectively.
|