Device Problem
Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.(b)(6).
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a power (intermittent power) issue.There was no additional information available for this complaint.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
|
|
Manufacturer Narrative
|
Device evaluation: the device has been returned and evaluated by product analysis on 05/10/2016 with the following findings: there was no damage found to the battery compartment or battery cap.The battery cap height and width contact measurements were within specification.The battery cap secured tightly to the pump and maintained electrical connection.During testing, the pump was powered on with auditory and vibratory tone; however, the display screen was blank.The remaining steps were unable to be adequately investigated for the reported ¿intermittent power¿ complaint.The pump¿s cover was removed; a single component failure was found on the printed circuit board (pcb).The pump was able to power on and displayed the verify screen after replacing the component on the pcb.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(4).
|
|