Model Number H749518080
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/26/2016
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: the device was returned for analysis.Device analysis revealed an open hole in the lapjoint area.Fluid was leaking from the lapjoint area when the catheter was flushed.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Event Description
Reportable based on device analysis completed on (b)(4) 2016.It was reported that shaft kinked occurred.The 90% stenosed target lesion was located in the moderately tortuous and moderately calcified distal circumflex artery (lcx).An opticross imaging catheter was selected for use.During preliminary intravascular ultrasound, the opticross imaging catheter was introduced; however, it was noticed that the shaft was kinked.The procedure was completed with another opticross imaging catheter.No patient complications were reported and the patient's status is good.However, device analysis revealed an open hole in the lap joint area.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS¿
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5570210
MDR Text Key42439198
Report Number2134265-2016-02601
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
PMA/PMN Number
K123621
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 03/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/10/2016
Device Model NumberH749518080
Device Catalogue Number51808
Device Lot Number18602823
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-