The cause for the falsely depressed total hcg result is unknown.The technical application specialist (tas) performed a precision run and the overall precision with the quality control (qc) and the patient samples yielded acceptable results.Siemens healthcare diagnostics is investigating.The ifu states in the limitations section: "test results alone are not diagnoses of medical conditions.For example, low titer elevations of hcg can occur in normal nonpregnant subjects.A physician's diagnosis involves evaluation of the test result in conjunction with, and in the context of, the patient's medical history, physical examination, and other test results, sometimes in consultation with other medical experts.".
|
Siemens filed the initial mdr 1219913-2016-00071 on april 21, 2016.On 04/25/2016 additional information: the patient sample was not available for in-house testing and investigation.The precision was checked by the technical application specialist (tas) at the customer site and no issues were identified.The precision was reviewed for in-house data and no imprecision was noted.The cause appears to be related to preanalytical handling including possibly fibrin, incomplete clotting and/ or incorrect spinning.The instrument is performing within specifications.No further evaluation of the device is required.
|