Model Number H7493924215050
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr.: device was returned for analysis.Microscopic and tactile examination revealed numerous kinks in the hypotube.Microscopic examination revealed a full separation at the collar/proximal location, with additional damage to the collar.Microscopic and tactile examination revealed the ptfe (polytetrafluoroethylene) was completely pulled out from the collar.The inner diameter was measured at the distal tip and found to be within specifications.Microscopic inspection found no irregularities or defects on the tip.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was reported that the catheter was defective.A guidezilla was selected for use.It was noted that the catheter was defective.No patient complications were reported.However, device analysis revealed that the shaft was broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEZILLA¿
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5597843
MDR Text Key43326240
Report Number2134265-2016-03512
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2017
Device Model NumberH7493924215050
Device Catalogue Number39242-1505
Device Lot Number18336834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/23/2016
Initial Date FDA Received04/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-