Model Number 0292
Device Problems Over-Sensing (1438); Pocket Stimulation (1463); Device Dislodged or Dislocated (2923)
Patient Problem Muscle Stimulation (1412)
Event Date 02/18/2016
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that the patient presented to the clinic with complaints of diaphragmatic stimulation.When attempting to program around the stimulation, crosstalk of the atrial pace on the ventricular channel was seen.It was noted that when the sensitivity of the right ventricular (rv) lead was reprogrammed to 1.0 millivolts the crosstalk disappeared.An x-ray was taken which revealed a micro-dislodgement.Subsequently a revision procedure was performed where the rv lead was successfully repositioned.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5601077
MDR Text Key43441645
Report Number2124215-2016-03969
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910073/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 02/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/08/2017
Device Model Number0292
Other Device ID NumberRELIANCE 4-SITE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 02/18/2016
Initial Date FDA Received04/22/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age68 YR
-
-