MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Device Problems Disconnection (1171); Leak/Splash (1354); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/30/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This is a report of a use error where the patient accidentally became disconnected during therapy.Use errors and proper user instructions are addressed in ¿the homechoice and homechoice pro systems patient at-home guide¿, which is shipped with every homechoice device.The guide instructs the user to check all disposable set connections for a secure fit before beginning therapy.Also, it provides step-by-step instructions for properly connecting the patient line to the transfer set.A review of the label for the product family will be conducted.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a patient line became disconnected and leaked onto the bed.This occurred during drain one of peritoneal dialysis therapy.During troubleshooting, it was determined that the patient accidentally became disconnected and the patient reconnected and continued therapy.The technical service representative assisted the patient with ending therapy.Proper procedures were discussed per the user manual.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5601495
MDR Text Key43449688
Report Number1416980-2016-07857
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 04/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Date Manufacturer Received03/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age71 YR
-
-