Technical operations investigated this issue through code review and patient data; and determined that the issue with the blood pressure measurement is associated with a software defect.When the software defect occurs, the patient's submitted measurements are not populated in the patient's data table.The clinician is unable to view the patient's measurements in the ecare coordinator clinical user interface.The customer does receive adherence flags indicating the patient's measurements have not been received.A correction to the software defect has been developed, reviewed and approved.The correction is released and philips visicu is working with customers to implement.
|