• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE FLORIDA OPERATIONS BED EXTENDER 9153650579; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA OPERATIONS BED EXTENDER 9153650579; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number IHSC900DLX
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.Reference report number 1031452-2016-01663 for serious injury previously reported.
 
Event Description
The dealer originally contacted us regarding a serious injury alleged, and after further inspection of an audit across all of their homes, they noticed that this edge was on all of the beds was ranging from a little roughness on the edge to a razor sharp corner.They had to have their maintenance directors go around and file each one down to avoid any future incidents.
 
Manufacturer Narrative
(b)(4).No immediate risk noted with event to escalate to capa/crt.One serious injury noted out of all (b)(4) bed throughout a single business unit.Complaint data over a 2 year period did not show any similar issues or injuries with the (b)(4) bed.Conclusions of erb show that the underlying cause appears to be metal burs left after molding.Should additional information become available a supplemental record will be filed.Reference report number 1031452-2016-01663 for serious injury previously reported.
 
Event Description
The dealer originally contacted us regarding a serious injury alleged, and after further inspection of an audit across all of their homes, they noticed that this edge was on all of the beds was ranging from a little roughness on the edge to a razor sharp corner.They had to have their maintenance directors go around and file each one down to avoid any future incidents.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BED EXTENDER 9153650579
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5607771
MDR Text Key44352976
Report Number1031452-2016-01818
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date 06/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberIHSC900DLX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-