Device evaluation summary: belt sn (b)(4) was returned to the distributor where it was scrapped due to a biohazard.The device is not available for further evaluation.Review of the available download data does not indicate any device malfunction contributing to the event.Monitor sn (b)(4) was returned to zmc where it passed incoming functional testing.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the device, followed by a 5hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the audio messaging and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.There is no evidence to suggest that the lifevest caused or contributed to the patient's death.The patient was in svt when the device was shutdown.The patient had been responding appropriately to the false detections prior to the device being shutdown.
|