MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RE OX W/ RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RE OX W/ RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25RE
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 04/11/2016
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular systems that prior to cardiopulmonary bypass, while removing the units from the box, it was noticed that the port off on the reservoir is damaged.No patient involvement as the event occurred out of box.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on april 29, 2016.The sample was not returned for evaluation.A retention sample with the same product code/lot number combination was obtained.Visual inspection was performed on the unit and it was confirmed to not have any damage or anomalies anywhere on the device.Review of the device history records revealed no manufacturing issues.A definitive root cause could not be determined; therefore, this complaint is not confirmed.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25RE OX W/ RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball rd
elkton MD 21921
Manufacturer Contact
shari bailey
125 blue ball rd
elkton, MD 21921
8002837866
MDR Report Key5620841
MDR Text Key44129726
Report Number1124841-2016-00185
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Model Number3CX*FX25RE
Device Catalogue NumberN/A
Device Lot NumberTM26
Other Device ID Number(01)00699753450462
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2016
Initial Date FDA Received04/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/29/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-