MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3365-40Q
Device Problem Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/31/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a telemetry anomaly was observed on the device during implant.Inductive telemetry was used following device implant but was unsuccessful in interrogating the device.Device was explanted and replaced.Patient was stable post-procedure.
 
Manufacturer Narrative
No conclusion code available.The reported inability to interrogate was confirmed in the laboratory.The rf telemetry was noted to be anomalous and was due to the rf module.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key5621418
MDR Text Key44117411
Report Number2938836-2016-03333
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/31/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2018
Device Model NumberCD3365-40Q
Device Lot NumberA000016854
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/31/2016
Initial Date FDA Received04/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received06/28/2016
08/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
-
-