Returned product consisted of the watchman access sheath (was) and dilator, which was inside the was as recieved.The dilator was removed during product analysis.The valve was opened as received with the dilator inside the was.Microscope examination of the valve did not reveal any damage or irregularities; however, it was noted that the threads of the was were damaged/cross threaded.It could not be determined when the thread damage occurred.Functional testing of the valve confirmed the ability to completely close the valve with minimal forward force after removing the dilator.Water was injected into the was by attaching a syringe filled with water.There were no leaks or air bubbles observed during water injection.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|