• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC EQUINOXE REVERSE SHOULDER COMPRESSION SCREW/LOCKING CAP KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC EQUINOXE REVERSE SHOULDER COMPRESSION SCREW/LOCKING CAP KIT Back to Search Results
Catalog Number 320-20-30
Device Problems Failure To Adhere Or Bond (1031); Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/13/2016
Event Type  malfunction  
Manufacturer Narrative
Engineering evaluation noted that the revision reported was likely the result of glenoid loosening related to a combination of the patient's age, bone quality, and implant placement.
 
Event Description
Index surgery: (b)(6) 2010.Revision of left shoulder components due to a loose glenosphere with a broken screw.Infection likely.Surgeon states event is unlikely related to devices.This event report was received through clinical data collection activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE SHOULDER COMPRESSION SCREW/LOCKING CAP KIT
Type of Device
COMPRESSION SCREW/LOCKING CAP KIT
Manufacturer (Section D)
EXACTECH, INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
graham cuthbert
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key5628425
MDR Text Key44587675
Report Number1038671-2016-00314
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Catalogue Number320-20-30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received01/27/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age88 YR
Patient Weight65
-
-