MAUDE Adverse Event Report: BAXTER HEALTHCARE CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 2C6255
Device Problems Kinked (1339); Failure to Prime (1492)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.During visual inspection, the tubing was kinked.The reported condition was verified.The cause of the condition could not be determined.An ncr has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink stopcock and i.V.Solution administration set had kinks that impeded fluid flow during priming.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO STOPCOCK AND I.V. SOLUTION ADMINISTRATION SETS
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE
600 mts oeste de entrada
cartago
CS 
Manufacturer (Section G)
BAXTER HEALTHCARE
600 mts oeste de entrada
cartago
CS  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5628746
MDR Text Key44313793
Report Number1416980-2016-08379
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130245
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date06/15/2020
Device Catalogue Number2C6255
Device Lot NumberR15F15115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/12/2016
Initial Date FDA Received05/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-