Model Number D012
Device Problems Failure to Capture (1081); Pocket Stimulation (1463)
Patient Problem Muscle Stimulation (1412)
Event Date 03/11/2016
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this device and right ventricular (rv) lead exhibited loss of capture and diaphragmatic stimulation.Subsequently, the rv lead was repositioned.This device remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5634514
MDR Text Key44506309
Report Number2124215-2016-07488
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526533822
UDI-Public(01)00802526533822(17)20160526
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960040/S306
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/26/2016
Device Model NumberD012
Other Device ID NumberINOGEN MINI ICD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 03/11/2016
Initial Date FDA Received05/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
D012
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age66 YR
-
-