• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER, INC. ACETABULAR SYSTEM FLEXIBLE DRIVER; HWR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER, INC. ACETABULAR SYSTEM FLEXIBLE DRIVER; HWR Back to Search Results
Model Number N/A
Device Problems Device Operates Differently Than Expected (2913); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/11/2015
Event Type  malfunction  
Manufacturer Narrative
Device was returned with complaint for an evaluation.A visual inspection of the returned flexible driver shows a deformed flex shaft.Device shows evidence of wear since lot number was worn and not legible for determination; therefore, a device history record review was not performed as the lot number was unknown.The device was used for treatment.The reported device has a field life of more than 15 years to date based on the markings on the device.This type of event may be related (but not limited) to: excessive force that may have been applied during usage, continued operation of the drill after it was stuck against hard material, rapid forward and reversal of direction of rotation when the device is stuck, excessive bending around corners, device wear due to usage with time, and low speed/high torque of operation.A definitive root cause for this event was not determined with the information provided.This mdr was identified during an internal retrospective review to meet reporting requirements and therefore is being filed retrospectively.
 
Event Description
It was reported that during surgery the flexible shaft instrument bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR SYSTEM FLEXIBLE DRIVER
Type of Device
HWR
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
carrie schneider
p.o. box 708
warsaw, IN 46581-0708
8006136131
MDR Report Key5637017
MDR Text Key44611670
Report Number0001822565-2016-01483
Device Sequence Number1
Product Code HWR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 03/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number36600066
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-