EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHI; SWAN-GANZ CATHETER Back to Search Results
Model Number 777HF8J
Device Problem Low Readings (2460)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2016
Event Type  malfunction  
Manufacturer Narrative
No product was returned for evaluation; it was discarded at the hospital.Without the return of the product, it is not possible to determine if damages or defects existed on the product.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.No actions will be taken at this time.
 
Event Description
It was reported that the swan ganz catheter was placed in the patient after cardiac surgery.After two to three days of use, cci value gradually decreased from about 2.5l/min/m2 to 1.5l/min/m2.The expected cci value was more than 2.5l/min/m2.The catheter position was checked and the catheter was in the correct position.The patient was not treated based on the incorrect value.Error message was not observed.Vig2 monitor and the cable were tested by the customer but there were no problems observed.No further information was able to be obtained.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER WITH AMC THROMBOSHI
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key5637482
MDR Text Key45315613
Report Number2015691-2016-01509
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number777HF8J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2016
Initial Date FDA Received05/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-