SYNCARDIA SYSTEMS, INC. SYNCARDIA FREEDOM AC POWER SUPPLY Back to Search Results
Catalog Number 295050-001
Device Problems Failure to Charge (1085); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 03/14/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom ac power supply was not charging the freedom onboard batteries in the freedom driver that was supporting a patient.The customer also reported that the patient was provided with a replacement ac power supply.There was no reported adverse patient impact.This alleged failure mode poses a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver continued to function as intended.The freedom driver has a redundant power source of onboard batteries.The freedom ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Manufacturer Narrative
(b)(4) follow-up.
 
Event Description
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer reported that the freedom ac power supply was not charging the freedom onboard batteries in the freedom driver that was supporting a patient.The customer also reported that the patient was provided with a replacement ac power supply.There was no reported adverse patient impact.The freedom ac power supply was returned to syncardia for evaluation.Visual inspection of the freedom ac power supply revealed a damaged strain relief, cracked connector cover, and damaged housing.The freedom ac power supply was functionally tested and failed to provide output voltage at the power adaptor connector.Further investigation revealed a broken inductor wire in the power supply and was therefore the root cause for the inability of the ac power supply to provide power to the freedom driver and charge the onboard batteries.The freedom ac power supply was taken out of service.This failure mode poses a low risk to the patient because it did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver continued to function as intended.Patients are also provided additional onboard batteries.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, INC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
carole marcot
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key5642479
MDR Text Key45888711
Report Number3003761017-2016-00187
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 03/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295050-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/14/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age23 YR
-
-