Device evaluated by mfr: returned product consisted of the watchman access sheath (was).The valve was opened, as received, with blood was on the device.The valve, hub, shaft, and tip were microscopically, tactile and visually inspected.Inspection revealed kinks in the sheath at 4.5cm, 45 cm and 64 cm from the tip, sheath damage (flattened) for 8 cm, tip damage (split), and hub damage (cross threading).Functional testing of the valve confirmed the ability to completely close the valve with forward force.Water was injected into the was by attaching a syringe filled with water.There were no leaks or air bubbles observed during water injection.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|