• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA®; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA®; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-1132
Device Problem Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
A report was received that the patient's ipg was not communicating with the remote control (rc).The patient underwent a revision procedure wherein the ipg was replaced.The patient was reportedly doing well postoperatively.
 
Manufacturer Narrative
Device evaluation indicated that the complaint was confirmed.Upon receiving, the device was in hibernation, and it was charged to 3.987 volts in one charging cycle.However, the charge and system data logs were corrupted, and the measurement of the battery internal resistance supported the battery tab anomaly.It was due to the intermittent connection to the battery tab and it is the source of the charging and communication anomalies.
 
Event Description
A report was received that the patient's ipg was not communicating with the remote control (rc).The patient underwent a revision procedure wherein the ipg was replaced.The patient was reportedly doing well postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA®
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key5658532
MDR Text Key45278031
Report Number3006630150-2016-01251
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/18/2017
Device Model NumberSC-1132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/20/2016
Initial Date FDA Received05/16/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
-
-