• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number 518-062
Device Problems Pacing Problem (1439); Retraction Problem (1536); Cut In Material (2454); Device Operates Differently Than Expected (2913)
Patient Problems Hemorrhage/Bleeding (1888); Perforation (2001); Tissue Damage (2104); Low Pulmonary Arterial Wedge Pressure (2479); Pericardial Effusion (3271)
Event Date 04/22/2016
Event Type  Injury  
Event Description
This was a procedure to replace a right atrial lead due to its inability to pace effectively.Once the pacer pocket was opened, it was appreciated that the insulation on the ra lead was severely compromised.The ra leads helix was retracted, the terminal pic cut off and an ez lld was place to the tip.A 12 fr glide light laser sheath was used to ablate the scar between the ra and rv leads to the right atrium.Once the laser was advanced to the ra the sheath was advanced to the ra tip.The physician attempted to dislodge the ra tip with appropriate counter traction without release of the lead.Due to this the physician requested the ct surgeon for consultation.Counter traction was again applied to the ra lead, which released it from the heart.Approximately five minutes after the lead extraction the patient arterial pressure was sagging and in the 60's.A tee showed a pericardial effusion, the ct surgeon returned and performed a pericardiocentesis which returned approximately 80 cc's of blood.The patient's pressure stabilized but failed to return to normal.The surgeon performed a sternotomy to repair an injury to the patients ra.It was sutured closed.The injury was reported to be round holes which appeared to be the size of the ra lead tip.It was theorized by the physician that the lead had perforated the ra at time of implant, 62 months earlier.The patient survived the procedure and was taken to icu.Her current status is good.This report is against the lld as it was the mechanism of traction.
 
Manufacturer Narrative
Device 510k number has been corrected to reflect the most current and up to date number, as of the date of the initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
SPECTRANETICS
9965 federal dr.
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key5659736
MDR Text Key45312174
Report Number1721279-2016-00064
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/23/2016
Initial Date FDA Received05/17/2016
Supplement Dates Manufacturer ReceivedNot provided
07/24/2017
Supplement Dates FDA Received03/23/2017
08/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age67 YR
Patient Weight62
-
-