• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ORTHO-CLINICAL DIAGNOSTICS VITROS CHEMISTRY PRODUCTS PHYT SLIDES; IN VITRO DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORTHO-CLINICAL DIAGNOSTICS VITROS CHEMISTRY PRODUCTS PHYT SLIDES; IN VITRO DIAGNOSTIC Back to Search Results
Catalog Number 8298671
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/19/2016
Event Type  malfunction  
Manufacturer Narrative
The investigation determined that multiple non-reproducible, lower than expected vitros phyt results were obtained from multiple patient samples on two vitros 5600 integrated systems.A likely assignable cause for the lower than expected vitros phyt results could not be determined, however, an unexpected instrument, a reagent issue, or user error due to inappropriate pre-analytical sample handling cannot be ruled out as contributing factors.The within run phyt precision testing was unacceptable, indicating the instrument was not performing as intended, however repeat within run phyt testing was acceptable without changing the state of the analyzer.Therefore the vitros 5600 system contribution to the event could not be ruled out or confirmed.Based on review of historical quality control, there was no indication the vitros phyt microslide reagent lot 2614-0158-8794 malfunctioned.No quality control results were provided for the alternate reagent lot 2614-0158-8775, used on the vitros j2 system, therefore, although there is no evidence the reagent lot malfunctioned, it cannot be ruled out as a contributing factor to the event.Pre-analytical sample processing could not be ruled out as a contributing factor, as the customer was unable to provide information regarding the specific centrifugation conditions used to pre-analytically handle the patient sample in question, therefore it is unknown if they are following the sample device manufacturer¿s recommended guidelines for sample centrifugation.Cellular debris, due to poor sample preparation, may have been present in the affected sample, although this could not be confirmed.
 
Event Description
A customer obtained multiple non-reproducible, lower than expected vitros phenytoin (phyt) results from multiple patient samples processed using vitros chemistry products phyt slides on two vitros 5600 integrated systems (j1 and j2).J1: vitros phyt reagent lot 2615-0158-8794.Patient a: vitros result 3.4 and 4.6 ug/ml vs.The expected result of 10.9 ug/ml.Patient c: vitros result 6.1 ug/ml vs.The expected result of 12.9ug/ml.Patient 2: vitros result 11.0 ug/ml vs.The expected result of 16.3 ug/ml.Patient 6: vitros result 3.3 and 8.2ug/ml vs.The expected result of 10.9 ug/ml.Patient 7: vitros result 16.2 and 23.4 ug/ml vs.The expected result of 35.7ug/ml.Patient 8: vitros result 3.0 and 3.0 ug/ml vs.The expected result of 24.7ug/ml.Patient 9: vitros result 12.5 ug/ml vs.The expected result of 16.3ug/ml.Patient 13: vitros result 3.0 ug/ml vs.The expected result of 7.5ug/ml.Patient 14: vitros result 3.0 ug/ml vs.The expected result of 7.5ug/ml.Patient 15: vitros result 9.4ug/ml vs.The expected result of 25.2 ug/ml.J2: vitros phyt reagent lot 2615-0158-8775.Patient 7: vitros result 27.7 ug/ml vs.The expected result of 35.7ug/ml.Patient 8: vitros result 3.0 and 3.0 ug/ml vs.The expected result of 24.7ug/ml.Patient 9: vitros result 11.3 ug/ml vs.The expected result of 16.3ug/ml.Patient 13: vitros result 3.0 ug/ml vs.The expected result of 7.5ug/ml.Patient 15: vitros result 5.3 ug/ml vs.The expected result of 25.2 ug/ml.The customer reported the result for patient a (3.4 ug/ml) to the clinician; however, the higher than expected result was questioned by the clinician and the corrected report was released.No additional results were reported from the laboratory, as the additional patient samples were tested for investigation purposes only.Ortho has not been made aware of any allegations of actual patient harm occurring as a result of the events.This report is number nine of twenty mdr's for this event.(b)(4) 3500a forms are being submitted for this event as (b)(4) devices were involved.This report corresponds to ortho clinical diagnostics inc.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS CHEMISTRY PRODUCTS PHYT SLIDES
Type of Device
IN VITRO DIAGNOSTIC
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS
100 indigo creek drive
rochester NY 14626
Manufacturer (Section G)
ORTHO-CLINICAL DIAGNOSTICS
513 technology blvd.
rochester NY 14652
Manufacturer Contact
james a stevens
100 indigo creek drive
rochester, NY 14626
5854533000
MDR Report Key5664722
MDR Text Key46666547
Report Number1319809-2016-00065
Device Sequence Number1
Product Code DIP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 05/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/15/2016
Device Catalogue Number8298671
Device Lot Number2615-0158-8794
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/19/2016
Initial Date FDA Received05/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-