The investigation determined that a lower than expected vitros phyt result was obtained from a single patient sample on a vitros 4600 chemistry system.The investigation was unable to determine a definitive assignable cause.Based on historical phyt quality control performance, a reagent issue is not a likely contributing factor to this event.A vitros 4600 chemistry system issue could not be ruled out as the customer declined precision testing to confirm instrument performance.In addition, improper pre-analytical sample processing could not be ruled out as a contributing factor as the customer was not adhering to the sample collection device manufacturer¿s recommendations for sample centrifugation.The assignable cause is unknown.
|
A customer obtained a lower than expected vitros phyt result from a single patient sample, using vitros phyt slides on a vitros 4600 chemistry system.Patient result = 8.2 ug/ml versus an expected result of 26.0 ug/ml.Biased results of the direction and magnitude observed could lead to inappropriate physician action.The lower than expected vitros phyt patient result was reported outside of the laboratory, however, the result was questioned by the physician and a repeat test was performed.A corrected report was subsequently issued and there was no allegation of actual patient harm as a result of this event.(b)(4).
|