• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3365-40Q
Device Problems Difficult to Interrogate (1331); Inappropriate or Unexpected Reset (2959)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).The reported field events of backup vvi and an inability to communicate with the device were confirmed in the laboratory.The device was tested on the bench and an anomalous component on the hybrid was found.The cause of the reported backup vvi and inability to communicate was an anomalous component on the hybrid.(b)(4).
 
Event Description
It was reported that the asymptomatic patient presented in the clinic for a scheduled device check-up.Upon device interrogation, the device could not be interrogated and was found to be in backup vvi mode.The device was explanted and replaced.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jaime chavez
645 almanor avenue
sunnyvale, CA 94085
8184934022
MDR Report Key5670820
MDR Text Key45643559
Report Number2938836-2016-04009
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2017
Device Model NumberCD3365-40Q
Device Lot NumberA000006213
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/05/2016
Initial Date FDA Received05/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
-
-