• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2016
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable free thyroxine (ft4) results for one patient sample from a cobas e602 analyzer.The sample was submitted for investigation and tested on an analytical e module (e170) and a cobas e411 analyzer.The customer's results were reported outside the laboratory.The patient was not adversely affected.An issue with the analyzer was excluded as the repeat results were comparable to the customer's results.
 
Manufacturer Narrative
A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.Possible assay or system specific reasons for the issue include bubbles or foam on reagent surface, bubbles or foam on the system reagent surface, and possibly old pinch tubes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5676898
MDR Text Key45857465
Report Number1823260-2016-00688
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/16/2016
Initial Date FDA Received05/24/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-