TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
|
Back to Search Results |
|
Model Number 164267 |
Device Problems
Break (1069); Component Missing (2306); Naturally Worn (2988)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The fsr replaced the parts (latch, spring and retainer) and all are now operating properly.The unit operated to manufacturer specifications and was returned to clinical use.The suspect parts were returned to the manufacturer, no further action is needed at this time.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, he found all three parts that hold the disposable on to the centrifugal drive motor are broken, cracked, or missing.There was no patient involvement.
|
|