W.L. GORE & ASSOCIATES GORE® ACUSEAL VASCULAR GRAFT; PROSTHESIS, VASCULAR GRAFT Back to Search Results
Catalog Number ECH470045A
Device Problems Material Frayed (1262); Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/12/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
On (b)(6) 2016, a gore acuseal vascular graft was implanted in patient's right thigh in an arteriovenous access procedure.On (b)(6) 2016, the patient presented for a declot procedure.A trerotola thrombectomy device was used in an attempt to declot the graft.The device was placed beyond the clot, and as the thrombectomy device was moved back and forth, sections of the inner walls of the graft was torn off.The frayed graft wall was pushed back into place with an implant of a gore viabahn endoprosthesis.The patient did not experience any adverse consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® ACUSEAL VASCULAR GRAFT
Type of Device
PROSTHESIS, VASCULAR GRAFT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
MEDICAL WEST B/P
1505 n. fourth street
flagstaff AZ 86004
Manufacturer Contact
genevieve begay
1500 n. 4th street
flagstaff, AZ 
9285263030
MDR Report Key5682900
MDR Text Key46075867
Report Number2017233-2016-00528
Device Sequence Number1
Product Code DSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/19/2018
Device Catalogue NumberECH470045A
Device Lot Number4836964PP018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
-
-