MAUDE Adverse Event Report: HILL-ROM MEXICO CAREASSIST BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO CAREASSIST BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 1170
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2016
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the brake microswitch out of adjustment.Per the hill-rom service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Plug the bed into an appropriate power source and set the brakes to neutral.Make sure the alarm is heard.Set the brakes on the bed.Make sure the alarm stops sounding.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2016.It is unknown if the facility performed any other preventative maintenance on this bed.The technician adjusted the brake microswitch to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the brake not set alarm was not working.The bed was located at the (b)(6).There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREASSIST BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
812931-286
MDR Report Key5688543
MDR Text Key46202177
Report Number3006697241-2016-00067
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model Number1170
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-