Model Number 37612
Device Problem Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/30/2016
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical devices: product id: 37612, serial# (b)(4), implanted: (b)(6) 2016, product type: implantable neurostimulator.
 
Event Description
Information was received from a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for dystonia and movement disorders.It was reported that following an implantable neurostimulator (ins) replacement, the low impedances that were noted prior to the replacement continued to occur; the bipolar pair 0 + 3 had an impedance of 37ohms while all other pairs were within the normal rage.It was noted that the cause of the impedances was unknown.It was confirmed that the connection to the battery was ok and the patient was receiving effective therapy.No symptoms were reported.
 
Event Description
Additional information received reported the short had first been noticed on (b)(6) 2016 when the implantable neurostimulator (ins) was interrogated during surgery.The original battery had displayed the same short.It was confirmed the patient was receiving sufficient therapy with the short so the new ins was implanted.No further troubleshooting was performed upon replacing the ins.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key5690820
MDR Text Key46266208
Report Number3004209178-2016-10627
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2016
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/11/2016
Initial Date FDA Received06/01/2016
Supplement Dates Manufacturer ReceivedNot provided
06/06/2016
Supplement Dates FDA Received06/30/2016
09/22/2017
Date Device Manufactured11/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
-
-