Catalog Number 2950000000 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/02/2016 |
Event Type
malfunction
|
Event Description
|
It was reported that the mattress had a burned plug receptacle.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
Supplemental submitted to indicate that the unit was repaired and returned.The investigation concluded that a likely cause for the burn marks could have been due to an overload situation or high current spikes.
|
|
Event Description
|
It was reported that the mattress had a burned plug receptacle.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|