• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO XPRT SLEEP SURFACE; BED, PATIENT ROTATION, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO XPRT SLEEP SURFACE; BED, PATIENT ROTATION, POWERED Back to Search Results
Catalog Number 2950000000
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2016
Event Type  malfunction  
Event Description
It was reported that the mattress had a burned plug receptacle.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
Supplemental submitted to indicate that the unit was repaired and returned.The investigation concluded that a likely cause for the burn marks could have been due to an overload situation or high current spikes.
 
Event Description
It was reported that the mattress had a burned plug receptacle.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPRT SLEEP SURFACE
Type of Device
BED, PATIENT ROTATION, POWERED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
mary klaver
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key5691136
MDR Text Key46288793
Report Number0001831750-2016-00186
Device Sequence Number1
Product Code IKZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number2950000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/02/2016
Initial Date FDA Received06/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/07/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-