MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/19/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A kink in the tubing during prime is a known cause of air in the patient line.As the cassette was not returned and the lot number was unknown, a device analysis could not be completed.The cause of the issue was determined to be due to kinked tubing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that air was observed in the patient line tubing during the initial drain of peritoneal dialysis therapy on the homechoice.The patient was connected at the time the air was observed.During troubleshooting, the patient reported that the patient line was kinked in the cassette door.The technical service representative assisted the patient in ending therapy and reviewed proper procedures with them.There was no patient injury or medical intervention reported.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5692589
MDR Text Key46363489
Report Number1416980-2016-10101
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 06/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Date Manufacturer Received05/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age48 YR
-
-