BIOMET ORTHOPEDICS G7 NEUTRAL ARCOMXL LNR 36MM G; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Component or Accessory Incompatibility (2897)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/29/2016
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under warnings, "improper selection, placement, positioning, alignment and/or fixation of the implant components may result in unusual stress conditions which may lead to subsequent reduction in the service life of the prosthetic components.".
 
Event Description
During a procedure, the liner would not seat properly and part of the inferior edge was not locking into the cup.Another liner was used to complete the procedure without significant delay.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.(b)(4).Examination of returned device found no evidence of product non-conformance.Review of the device was unable to confirm the reported condition.A conclusive root cause of the event could not be determined.Corrective action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LNR 36MM G
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key5704168
MDR Text Key46761913
Report Number0001825034-2016-01950
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/30/2020
Device Model NumberN/A
Device Catalogue Number010000742
Device Lot Number3588441
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/09/2016
Initial Date FDA Received06/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age60 YR
-
-