Model Number 1095060
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2016
Event Type  malfunction  
Event Description
Our customer reported this device to be defective.On return of the device for evaluation, it was noted that one prong was missing from the plug, and the other was bent.Our evaluation concluded that the plug may have been damaged by the end user applying excessive force, but there is a possibility that it broke off in the mains socket - if this was the case then a risk of exposure to a live part would be created.There is no report of any patient harm, and the device is not life-supporting or life-sustaining.The ifu for the device contains the warning 'the power supply cord cannot be replaced by the user.In case the power supply cord becomes damaged, contact philips respironics customer service for replacement (there is no service option)' internal reference ra 306457714.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ESSENCE
Type of Device
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
Manufacturer Contact
alessandro agosti
chichester business park
city fields way, tangmere
chichester, west sussex PO20 -2FT
UK   PO20 2FT
704231549
MDR Report Key5705420
MDR Text Key46777927
Report Number9681154-2016-00007
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042655
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 05/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number1095060
Device Catalogue Number1099966
Device Lot Number141209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2016
Initial Date Manufacturer Received 05/27/2016
Initial Date FDA Received06/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-