MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Unintended Arm Motion (1033); Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 05/16/2016
Event Type  malfunction  
Manufacturer Narrative
Return requested.Medtronic investigation of returned suspect device finds they were unable to tighten articulating arm handle to perform test.The arm handle was in its loose state and the handle was locked up.Mechanical failure - malfunction, locked-up handle hardware investigation was completed.This issue was found related to a hardware issue and was documented in a medtronic hardware anomaly tracking database.No further issues have been reported.Site declined to replace the damaged articulating arm.
 
Event Description
A medtronic representative reported a site navigation system articulating arm that would not tighten properly.No further details regarding the damage, or how it occurred, were provided.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
judith ericson
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key5707385
MDR Text Key46861910
Report Number1723170-2016-01060
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 06/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received05/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/19/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-