• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GERMANY ARCHITECT HE4

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GERMANY ARCHITECT HE4 Back to Search Results
Catalog Number 02P54-25
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The product was not available for return.An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications.Based on all available information and abbott diagnostics' complaint investigation, the assay performed as intended and no product deficiency was identified.
 
Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.(b)(4).
 
Event Description
The customer observed a false negative he4 result on the architect i1000sr analyzer.The following data was provided: initial 47.5, repeat 67.3 pmol/l.There was no impact to patient management reported.
 
Manufacturer Narrative
The conclusion code was corrected to (b)(4).The device evaluation was reassessed and concluded that a malfunction occurred, therefore, the device was not performing as intended, however, no systemic issue or product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT HE4
Type of Device
HE4
Manufacturer (Section D)
ABBOTT GERMANY
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GERMANY
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key5710065
MDR Text Key47944433
Report Number3002809144-2016-00035
Device Sequence Number1
Product Code OIU
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K093957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/05/2016
Device Catalogue Number02P54-25
Device Lot Number50440M500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/20/2016
Initial Date FDA Received06/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received06/21/2016
05/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age40 YR
-
-