Model Number 1095060
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/28/2016
Event Type  malfunction  
Event Description
Customer reported that this device had a bad odor, no airflow, and a possible loose internal component.On return of the device for evaluation it was apparent that the insulation on the mains cord was degraded to the extent that bare wires were visible.It is not clear how the degradation occurred, but the evaluation concluded that the device may have been damaged by the end user, as the power cord appeared to have been subjected to excessive abrasive forces.The ifu for this device contains the warning 'the power supply cord cannot be replaced by the user.In case the power supply cord becomes damaged, contact philips respironics customer service for replacement (there is no service option)'.No patient harm or death has been reported and the device is not life sustaining/supporting.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ESSENCE
Type of Device
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
Manufacturer Contact
alessandro agosti
chichester business park
city fields way, tangmere
chichester, west sussex PO20 -2FT
UK   PO20 2FT
704231549
MDR Report Key5716797
MDR Text Key47137101
Report Number9681154-2016-00006
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042655
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 05/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number1095060
Device Catalogue Number1100312
Device Lot Number150310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2016
Date Manufacturer Received05/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-