Results: the indigo system cat3 aspiration catheter (cat3) was kinked in multiple locations along the distal shaft.Conclusions: evaluation of the returned device revealed it was kinked in multiple locations along the distal shaft.This damage likely occurred due to forceful manipulation of the cat3 during advancement through tortuous patient anatomy.The loss of suction reported in the complaint likely occurred due to the kinks in the distal shaft of the cat3.Penumbra catheters are 100% visually inspected during in-process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure using an indigo system aspiration catheter 3 (cat3).During the procedure, the physician advanced the cat3 through the patient's tortuous anatomy into the pedal arch and then started removing clot from the patient.However, after three passes, the physician lost suction.Therefore, the cat3 was removed.After removal of the cat3, a kink was noticed at about 3-4 centimeters from the distal tip of the cat3.The physician indicated that the cat3 may have prolapsed on itself and kinked when he was advancing it.The procedure was completed using a new cat3.There was no report of an adverse effect to the patient.
|