TELEFLEX MEDICAL HUDSON NEBULIZER W/ADULT MASK & TBG, SMALL VOL Back to Search Results
Catalog Number 41893
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/25/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.No photo for review.The device history record (dhr) of batch number (b)(4) that belong to catalog number 41893 has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved in this complaint.However , material from the production line was inspected and no issues were detected that can lead this customer complaint.If the device sample becomes available at a later date this complaint will be re-opened.
 
Event Description
The customer alleges that there is a leak at one of the connections on the tubing.
 
Manufacturer Narrative
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed on the returned sample and the device performed properly.The device passed the leak test.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
 
Event Description
The customer alleges that there is a leak at one of the connections on the tubing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER W/ADULT MASK & TBG, SMALL VOL
Type of Device
NEBULIZER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
warrenda peterson
3015 carrington mill blvd
morrisville, NC 27560
9193613959
MDR Report Key5722775
MDR Text Key47313482
Report Number3004365956-2016-00242
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeSF
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number41893
Device Lot Number74H1503126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/03/2016
Initial Date FDA Received06/14/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-